![Agilent Technologies 5975C TAD VL MSD Скачать руководство пользователя страница 111](http://html1.mh-extra.com/html/agilent-technologies/5975c-tad-vl-msd/5975c-tad-vl-msd_troubleshooting-and-maintenance-manual_2867605111.webp)
General Maintenance
4
5975 Series MSD Troubleshooting and Maintenance Manual
111
To Purge the Calibration Valves
EI calibration valve
After adding new PFTBA to the calibrant vial, you must purge the air out of
the vial and valve.
1
If the vacuum gauge controller is on, turn it off.
2
In Tune and Vacuum Control view, select
Purge Calibrant Valve
under the
Vacuum menu.
This will open the CI calibration valve for several minutes with all analyzer
voltages turned off.
CI calibration valve
After adding new PFDTD to the calibrant vial, you must purge the air out of
the vial and valve.
1
If the vacuum gauge controller is on, turn it off.
2
Select
Gas A.
3
Verify that
PCICH4.U
is loaded.
4
In Diagnostics and Vacuum Control view, select
Purge Calibrant Valve
under
the Vacuum menu.
This will open the CI calibration valve for several minutes with all analyzer
voltages turned off.
C A U T I O N
After removing a calibrant vial, you
must
purge the calibration valve. Failure to do so
will result in damage to the filaments and the electron multiplier.
Содержание 5975C TAD VL MSD
Страница 1: ...Agilent Technologies Agilent 5975 Series MSD Troubleshooting and Maintenance Manual...
Страница 8: ...8 5975 Series MSD Troubleshooting and Maintenance Manual...
Страница 56: ...56 5975 Series MSD Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Страница 82: ...82 5975 Series MSD Troubleshooting and Maintenance Manual 3 CI Troubleshooting...
Страница 196: ...196 5975 Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System...
Страница 230: ...230 5975 Series MSD Troubleshooting and Maintenance Manual 8 Electronics...
Страница 249: ...Parts 9 5975 Series MSD Troubleshooting and Maintenance Manual 249 Figure 59 Turbo pump and related parts 1 2 3 4 5...
Страница 253: ...Parts 9 5975 Series MSD Troubleshooting and Maintenance Manual 253 Figure 61 Analyzer parts 7 under clips 3 4 5 6...
Страница 256: ...256 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 62 EI source 1 9 10 12 11 4 8 5 3 3 7 7 2...
Страница 270: ...270 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts...