7
Using 8480 Series Power Sensors
190
N1911A/1912A P-Series Power Meters User’s Guide
4
Press
to complete the entry.
Enter (or edit) the frequency and
cal factor pairs as follows:
1
Press
to add a new frequency value (or press
to edit).
Use the numeric keypad to enter the required value in the
Frequency
pop- up window. Complete the entry by pressing the
,
keys.
2
Enter the new cal factor value (or press
to edit). Use the
numeric keypad to enter the required value in the
Cal Factor
pop- up
window. Complete the entry by pressing the
key.
3
Continue adding/editing values until you have entered all the data you
require.
4
When you have finished editing the table press
to save the table.
Enter
N O T E
A calibration factor in the range of 1 % to 150 % can be entered.
The following rules apply to naming sensor calibration tables:
•
The name must consist of no more than 12 characters.
•
All characters must be upper or lower case alphabetic characters, or numeric (0-9), or
an underscore (_).
•
No other characters are allowed.
•
No spaces are allowed in the name.
Insert
Change
GHz MHz
Change
%
Done
N O T E
Ensure that the frequency points you use cover the frequency range of the signals you want
to measure. If you measure a signal with a frequency outside the frequency range defined
in the sensor calibration table, the power meter uses the highest or lowest frequency point
in the sensor calibration table to calculate the offset
Содержание 1912A
Страница 1: ...Agilent Technologies Agilent N1911A 1912A P Series Power Meters User s Guide...
Страница 9: ...N1911 1912A P Series Power Meters User s Guide ix Out position of bi stable push control...
Страница 24: ...xxiv N19121A 1912A P Series Power Meters User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 26: ...xxvi N19121A 1912A P Series Power Meters User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 278: ...10 Maintenance 252 N1911A 1912A P Series Power Meters User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...