Using E9300 E-Series Power Sensors
5
N1911A/1912A P-Series Power Meters User’s Guide
161
5
press
to display the
Meas Avg Count
pop- up.
6
Use the numeric keys to enter the required value.
7
Press
to complete the entry.
Procedure
Switch off step detection as follows:
1
Press
.
On dual channel meters, select the required channel.
2
Use the
and
keys to select the
Step Detect
setting field.
3
Press
to check the step detection to
Off
.
4
Press
key to close the
Channel Setup
screen.
Achieving Stable Results with GSM Signals
Signals with a pulse repetition frequency (PRF) close to a multiple or
sub- multiple of the 440 Hz chopper- amplifier signal generate a beat note
at a frequency between the PRF and 440 Hz. Control over the filter
settings is again required to obtain stable results.
Enter
N O T E
You should also ensure the filter is not reset when a step increase or decrease in power is
detected by switching the step detection off.
TIP
The PRF of a GSM signal is approximately 217 Hz and thus
requires more averaging than most other TDMA signals. To
achieve a stable measurement use the filter setting procedures
to set the
. Experimentally, a
setting of 148
gives optimum results although settings in the order of 31 or 32
give acceptable results if a faster measurement is required.
Length
Length
Содержание 1912A
Страница 1: ...Agilent Technologies Agilent N1911A 1912A P Series Power Meters User s Guide...
Страница 9: ...N1911 1912A P Series Power Meters User s Guide ix Out position of bi stable push control...
Страница 24: ...xxiv N19121A 1912A P Series Power Meters User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 26: ...xxvi N19121A 1912A P Series Power Meters User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 278: ...10 Maintenance 252 N1911A 1912A P Series Power Meters User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...