Specifications and Characteristics
11
N1911A/1912A P-Series Power Meters User’s Guide
263
1 mW Power Reference
N O T E
The
1mW Power Reference
is provided for calibration of the E-Series, 8480 Series and the
N8480 Series power sensors. The P-Series sensors are automatically calibrated, hence, do
not need this power reference for calibration.
Power Output:
1.00 mW (0.0 dBm)
Factory set to ±0.4 % traceable to the National Physical
Laboratories (NPL), UK
Accuracy:
(over 2-years)
±1.2 % (0 - 55 ºC)
±0.4 % (25 ± 10 ºC)
Frequency:
50 MHz nominal
SWR:
1.08 (0 - 55 ºC)
1.05 typical
Connector Type:
Type N (f), 50
W