Specifications and Characteristics
11
N1911A/1912A P-Series Power Meters User’s Guide
269
Noise and Drift
11
Within 1 hour after a zero, at a constant temperature, after 24 hour warm-up of the power meter.
This component can be disregarded with Auto-zero mode set to ON.
12
Measured over a one-minute interval, at a constant temperature, two standard deviations, with
averaging set to 1.
Effect of Video Bandwidth Setting:
The noise per sample is reduced by applying the meter video bandwidth filter setting (High,
Medium or Low). If averaging is implemented, this will dominate any effect of changing the
video bandwidth.
Sensor
Model
Zeroing
Zero Set
Zero
Drift
11
Noise per
Sample
Measurement
Noise
(Free Run)
12
<500 MHz
> 500 MHz
N1921A /
N1922A
No RF at
input
200 nW
100 nW
2 µW
50 nW
RF Present
550 nW
200 nW
Measurement
Average Setting
1
2
4
8
16
32
64
128
256
512
1024
Free Run Noise
Multiplier
1
0.9
0.8
0.7
0.6
0.5
0.45
0.4
0.3
0.25
0.2
Video BW setting
Low
5 MHz
Medium
15 MHz
High
30 MHz
Off
Noise per Sample
Multiplier
<
500 MHz
0.5
1
2
1
≥
500 MHz
0.45
0.75
1.1
1
Содержание 1912A
Страница 1: ...Agilent Technologies Agilent N1911A 1912A P Series Power Meters User s Guide...
Страница 9: ...N1911 1912A P Series Power Meters User s Guide ix Out position of bi stable push control...
Страница 24: ...xxiv N19121A 1912A P Series Power Meters User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 26: ...xxvi N19121A 1912A P Series Power Meters User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 278: ...10 Maintenance 252 N1911A 1912A P Series Power Meters User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...