Specifications and Characteristics
11
N1911A/1912A P-Series Power Meters User’s Guide
267
System Specifications and Characteristics
The video bandwidth in the meter can be set to High, Medium, Low and
Off. The video bandwidths stated in the table below are not the 3 dB
bandwidths, as the video bandwidths are corrected for optimal flatness
(except the Off filter). Refer to
Figure 11- 159
for information on the
flatness response of the High, Medium, Low and Off filters. The Off video
bandwidth setting provides the warranted rise time and fall time
specification and is the recommended setting for minimizing overshoot on
pulse signals.
Rise Time, Fall Time, and Overshoot versus Video Bandwidth
Settings
9
Specified as 10 % to 90 % for rise time and 90 % to 10 % for fall time on a 0 dBm pulse.
10
Specified as the overshoot relative to the settled pulse top power.
Parameter
Video Bandwidth Setting
Low:
5 MHz
Medium:
15 MHz
High:
30 MHz
Off
< 500 MHz
> 500 MHz
Rise Time/ Fall Time
9
<
56 ns
<
25 ns
≤
13 ns
<
36 ns
≤
13 ns
Overshoot
10
-
-
-
<
5 %
<
5 %
N O T E
For option 107 (10 m cable), add 5 ns to the rise time and fall time specifications.
Содержание 1912A
Страница 1: ...Agilent Technologies Agilent N1911A 1912A P Series Power Meters User s Guide...
Страница 9: ...N1911 1912A P Series Power Meters User s Guide ix Out position of bi stable push control...
Страница 24: ...xxiv N19121A 1912A P Series Power Meters User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 26: ...xxvi N19121A 1912A P Series Power Meters User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 278: ...10 Maintenance 252 N1911A 1912A P Series Power Meters User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...