Using E9320 E-Series Power Sensors
4
N1911A/1912A P-Series Power Meters User’s Guide
147
Improving the Accuracy with Additional Settings
You can now return to the setup menus and adjust the default setting to
obtain greater measurement accuracy.
Channel Setup
• Press .
The
Channel Setup
screen is displayed. If using an
N1912A, select the channel you want to configure.
Figure 4-111
E-Series E9320 power sensor dafault channel setup
Use the
,
,
, and
keys to highlight the setting field and press
to configure the setting. For further detail on these settings, refer
to the appropriate sections.
•
Meas Avg
:
“Setting Measurement Averaging”
on page 49
•
Video Avg
:
“Setting the Video Averaging”
on page 52
•
Step Detec
t:
“Step Detection”
on page 51
•
Video B/W
:
“Setting the Video Bandwidth”
on page 54
N O T E
For range settings, only UPPER and LOWER ranges are available in Normal and Triggered
mode.