10
Maintenance
234
N1911A/1912A P-Series Power Meters User’s Guide
Error Message List
–
101
Invalid character
An invalid character was found in the command string. You may have inserted a
character such as #, $, or % in the command header or within a parameter.
For example, LIM:LOW O#.
–
102
Syntax error
Invalid syntax was found in the command string.
For example, LIM:CLE:AUTO, 1 or LIM:CLE: AUTO 1.
–
103
Invalid separator
An invalid separator was found in the command string. You may have used a comma
instead of a colon, semicolon, or blank space; or you may have used a blank space
instead of a comma.
For example, OUTP:ROSC,1.
-105
GET not allowed
A Group Execute Trigger (GET) is not allowed within a command string.
–
108
Parameter not allowed
More parameters were received than expected for the command. You may have entered
an extra parameter, or added a parameter to a command that does not accept a
parameter.
For example, CAL 10.
–
109
Missing parameter
Fewer parameters were received than expected for the command. You omitted one or
more parameters that are required for this command. For example, AVER:COUN.
–
112
Program mnemonic too long
A command header was received which contained more than the maximum 12
characters allowed.
For example, SENSeAVERageCOUNt 8.
Содержание 1912A
Страница 1: ...Agilent Technologies Agilent N1911A 1912A P Series Power Meters User s Guide...
Страница 9: ...N1911 1912A P Series Power Meters User s Guide ix Out position of bi stable push control...
Страница 24: ...xxiv N19121A 1912A P Series Power Meters User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 26: ...xxvi N19121A 1912A P Series Power Meters User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 278: ...10 Maintenance 252 N1911A 1912A P Series Power Meters User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...