7
Using 8480 Series Power Sensors
188
N1911A/1912A P-Series Power Meters User’s Guide
Editing frequency dependent offset tables requires the following steps:
1
Identify and select the table you want to edit
2
Rename the table
3
Enter the frequency and offset pairs
4
Save the table
Procedure
First select the table you want to edit as follows:
1
Press
,
,
to display the
Sensor Tbls
screen.
Figure 7-131
“Sensor Tbls” screen
2
Choose the table you want to edit using the
and
keys. Press
to display the
Edit Cal
screen as shown in
Figure 7- 132
.
Tables
Sensor Cal Tables
Edit Table