General Power Meter Functions
2
N1911A/1912A P-Series Power Meters User’s Guide
87
Setting Limits
The power meter can be configured to verify the current measurement in
any measurement line against predefined upper and/or lower limit values.
The range of values that can be set for the upper and lower limits and the
default values depends on the measurement units in the currently selected
measurement line (see
Table 2- 7
).
Table 2-7
Range of values for window limits
Procedure
Set the limits as follows:
1
Press
,
to display the
Measurement Setup
menus.
2
Use the
and
keys to highlight the
Limits:
setting field.
3
Press
to check the
Limits:
setting field.
4
Use the
key to highlight the
Minimum Limits:
value field.
Window
Units
Maximum
Minimum
Default
Maximum
Default
Minimum
dB
+200 dB
–180 dB
60 dB
–120 dB
dBm
+230 dBm
–150 dBm
90 dBm
–90 dBm
%
999.9 X%
100.0 a%
100.0 M%
100.0 p%
W
100.000 XW
1.000 aW
1.000 MW
1.000 pW
N O T E
Ensure you have selected the channel you wish to set up.
Meas Select
Содержание 1912A
Страница 1: ...Agilent Technologies Agilent N1911A 1912A P Series Power Meters User s Guide...
Страница 9: ...N1911 1912A P Series Power Meters User s Guide ix Out position of bi stable push control...
Страница 24: ...xxiv N19121A 1912A P Series Power Meters User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 26: ...xxvi N19121A 1912A P Series Power Meters User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 278: ...10 Maintenance 252 N1911A 1912A P Series Power Meters User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...