General Power Meter Functions
78
N1911A/1912A P-Series Power Meters User’s Guide
17
Poll the status of the power meter by sending
*ESR?
.
*ESR?
will return
a 1 when buffering is completed. Use
FETCh?
to retrieve all the
buffered measurement.
Example of SCPI commands set
Refer to the following SCPI commands for normal and average mode in
power sweep measurement.
Normal mode
This feature is applicable for E9320 and N1920 Series power sensor when
connected. Below are examples of SCPI commands when used in normal
mode. Send the SCPI commands below to meter using remote interface.
•
SENS:DET:FUNC NORM
•
TRIG:SOUR EXT
•
SENS:BUFF:COUN 2048
Optional measurement configurations:
•
TRIG:DEL 1ms
•
TRIG:HOLD 10us
•
OUTP:TRIG ON
•
SENS:SWE1:OFFS:TIME 100us
•
SENS:SWE1:TIME 100us
•
INIT:CONT ON
•
*OPC
•
*ESR?
•
FETC?
N O T E
Refer to the
P-Series Power Meters Programming Guide
for the commands usage details.