Seconds Per Division Field
Use the sec/Div field to select the seconds per division resolution. The range
of the sec/Div field is 250 ps/Div to 50 s/Div. You set the sec/Div field by
either selecting the sec/Div field and rotating the knob, or by selecting the
sec/Div field twice, and then using the pop-up keypad. When using the knob
to set the sec/Div, the value will change in a 1-2-5 sequence. By adjusting the
sec/Div, you can zoom in to view a desired part of the display.
Seconds Per Division Field
Seconds per division
field
The Waveform Menu
Seconds Per Division Field
6–5
Содержание 16517A
Страница 3: ...The Agilent Technologies 16517A 4 GSa S Timing and 1 GSa S State Logic Analyzer ii ...
Страница 5: ...iv ...
Страница 11: ...Contents 6 ...
Страница 12: ...1 General Information ...
Страница 18: ...2 Probing ...
Страница 29: ...2 12 ...
Страница 30: ...3 The Format Menu ...
Страница 42: ...4 The Trigger Menu ...
Страница 79: ...4 38 ...
Страница 80: ...5 The Listing Menu ...
Страница 86: ...6 The Waveform Menu ...
Страница 106: ...7 The Compare Menu ...
Страница 122: ...8 The Chart Menu ...
Страница 133: ...8 12 ...
Страница 134: ...9 Markers and Stop Measurements ...
Страница 149: ...9 16 ...
Страница 150: ...10 Skew Adjust ...
Страница 156: ...11 Error Messages ...
Страница 165: ...11 10 ...
Страница 166: ...12 Specifications and Characteristics ...
Страница 176: ...13 Installation and Service ...
Страница 189: ...13 14 ...
Страница 193: ...Index 4 ...