X-O
When X-O is selected, a repetitive run is stopped when a comparison of the
time period between the X and O pattern markers and one of the following
time period options is true.
Less Than
X-O time must be less than the time value that you specify
in the Time field.
Greater Than
X-O time must be greater than the time value that you
specify in the Time field.
In Range
X-O time must be within the time range value that you specify
in the two Time fields.
Not in Range
X-O time must not be within the time range value that
you specify in the two Time fields.
X-O Stop Measurement Type Fields
X-O field
X-O options menu
The Compare Menu
Specify Stop Measurement Field
7–13
Содержание 16517A
Страница 3: ...The Agilent Technologies 16517A 4 GSa S Timing and 1 GSa S State Logic Analyzer ii ...
Страница 5: ...iv ...
Страница 11: ...Contents 6 ...
Страница 12: ...1 General Information ...
Страница 18: ...2 Probing ...
Страница 29: ...2 12 ...
Страница 30: ...3 The Format Menu ...
Страница 42: ...4 The Trigger Menu ...
Страница 79: ...4 38 ...
Страница 80: ...5 The Listing Menu ...
Страница 86: ...6 The Waveform Menu ...
Страница 106: ...7 The Compare Menu ...
Страница 122: ...8 The Chart Menu ...
Страница 133: ...8 12 ...
Страница 134: ...9 Markers and Stop Measurements ...
Страница 149: ...9 16 ...
Страница 150: ...10 Skew Adjust ...
Страница 156: ...11 Error Messages ...
Страница 165: ...11 10 ...
Страница 166: ...12 Specifications and Characteristics ...
Страница 176: ...13 Installation and Service ...
Страница 189: ...13 14 ...
Страница 193: ...Index 4 ...