Compare Full / Compare Partial Field
The Compare Full/Compare Partial field is a toggle field that allows you to
compare either the full range of states or define a subset of the total number
of states in the Reference image for use in the comparison.
The Compare mode is accessed by touching the Compare Full/Compare
Partial field in either the Compare or Difference listing menus. When
selected, a pop-up menu appears in which you select either the Full or Partial
option.
When you select the Partial option, fields appear for setting the start state
and stop state values. Only bits in states (lines) on or between the
boundaries are compared against the acquired data.
Compare Full / Compare Partial Field
Compare full / Compare
partial field
Partial range fields
The Compare Menu
Compare Full / Compare Partial Field
7–9
Содержание 16517A
Страница 3: ...The Agilent Technologies 16517A 4 GSa S Timing and 1 GSa S State Logic Analyzer ii ...
Страница 5: ...iv ...
Страница 11: ...Contents 6 ...
Страница 12: ...1 General Information ...
Страница 18: ...2 Probing ...
Страница 29: ...2 12 ...
Страница 30: ...3 The Format Menu ...
Страница 42: ...4 The Trigger Menu ...
Страница 79: ...4 38 ...
Страница 80: ...5 The Listing Menu ...
Страница 86: ...6 The Waveform Menu ...
Страница 106: ...7 The Compare Menu ...
Страница 122: ...8 The Chart Menu ...
Страница 133: ...8 12 ...
Страница 134: ...9 Markers and Stop Measurements ...
Страница 149: ...9 16 ...
Страница 150: ...10 Skew Adjust ...
Страница 156: ...11 Error Messages ...
Страница 165: ...11 10 ...
Страница 166: ...12 Specifications and Characteristics ...
Страница 176: ...13 Installation and Service ...
Страница 189: ...13 14 ...
Страница 193: ...Index 4 ...