![Agilent Technologies 16517A Скачать руководство пользователя страница 160](http://html1.mh-extra.com/html/agilent-technologies/16517a/16517a_user-reference_2867823160.webp)
Machine name: "filename" inverse assembler not found.
This
message appears when the inverse assembler file could not be found.
Need to have two free sequence levels.
For each sequence level
with a "<" assigned, you must leave two sequence levels free. To free up
two levels, simply delete two levels.
NOTE: 500 MHz Trigger Sequencer Rate Exceeds Sample Rate.
This message appears when the sample period in the acquisition control
menu is set to a lower rate than the sequencer. Because of the slow
sampling rate, the trigger may not be included in the trace.
NOTE: Sample Rate Exceeds 500 MHz Trigger Sequencer Rate.
This message appears when the sample period in the acquisition control
menu is set to a faster rate than the trigger sequencer.
NOTE: Trigger Sequencer Will Not See Oversampled States.
The
user has changed the oversampling value. This message warns that if
the trigger occurs on an oversampled state, the sequencer will not detect
it and thus will not trigger on that instance.
Pods have been truncated.
User is loading a configuration with
number of chips greater than the current system’s number of chips.
s/Div set to limit.
The sec/Div or states/Div field is set to its limit.
Search failed - X pattern not found.
The X pattern specified could
not be found, therefore the pattern marker could not be placed in the
data.
Search failed - O pattern not found.
The O pattern specified could
not be found, therefore the marker could not be placed in the data.
Skew adjustment halted, recalling stored values
The user has
selected Exit after beginning the skew adjust procedure.
Slow or missing clock.
This message indicates a slow or missing clock.
This is displayed for a state analyzer only until the first clock occurs.
The clock is not connected properly.
The module did not detect a
clock. A clock signal is necessary for skew adjustment.
Trigger May Not Be Included In Acquired Data.
The trigger event,
although detected, may not be included in the displayed trace because
the sample rate is low. You may want to decrease the sample period in
the acquisition control menu. See the chapter on the Trigger menu.
Error Messages
Warning Messages
11–5
Содержание 16517A
Страница 3: ...The Agilent Technologies 16517A 4 GSa S Timing and 1 GSa S State Logic Analyzer ii ...
Страница 5: ...iv ...
Страница 11: ...Contents 6 ...
Страница 12: ...1 General Information ...
Страница 18: ...2 Probing ...
Страница 29: ...2 12 ...
Страница 30: ...3 The Format Menu ...
Страница 42: ...4 The Trigger Menu ...
Страница 79: ...4 38 ...
Страница 80: ...5 The Listing Menu ...
Страница 86: ...6 The Waveform Menu ...
Страница 106: ...7 The Compare Menu ...
Страница 122: ...8 The Chart Menu ...
Страница 133: ...8 12 ...
Страница 134: ...9 Markers and Stop Measurements ...
Страница 149: ...9 16 ...
Страница 150: ...10 Skew Adjust ...
Страница 156: ...11 Error Messages ...
Страница 165: ...11 10 ...
Страница 166: ...12 Specifications and Characteristics ...
Страница 176: ...13 Installation and Service ...
Страница 189: ...13 14 ...
Страница 193: ...Index 4 ...