OsmoTECH XT Single-Sample Micro-Osmometer Service Guide
47
Chapter 7:
Troubleshooting
If a system error message appears on the screen during
operation and you cannot resolve the problem, contact
Advanced Instrument Technical Service for assistance
(see
Contacting Advanced Instruments
).
For quickest results, have the model and serial number
from the bottom of the instrument available when you
speak to a technician.
If you need to send your OsmoTECH XT for servicing, a
technical service representative will provide you with an
RMA number. Units are not accepted for service without
an RMA number.
CAUTION:
Do not open or remove the instrument
covers unless you are advised to do so by a service
engineer.
If you are advised to open or remove the
instrument covers, always unplug the power cord first.
You could be exposed to electric shock, excessive
temperatures, or mechanical hazards
7.1 General errors
The following table lists general errors that the system
can detect.
Error
Suggested action
AI consumable box
not detected
•
Place a test kit on the instrument.
Printer offline
•
Close printer door
•
Replace printer paper
•
Connect printer
7.2 Errors during calibration
For a list of errors that can occur during calibration, see
7.3 Errors during testing
If an error occurs during a test, a message displays on
the screen and is recorded in the events database.
Discard the sample, carefully clean the plunger, and then
try a new sample. If the error occurs again, refer to the
following table to help resolve the problem.
Error
Possible cause/Suggested action
Out of
range Low
Out of
range High
Cause:
The sample might be outside the range
of the instrument:
•
Out of range Low:
The measured reading is
below -10 mOsm.
•
Out of range High
:
–
The measured reading is above
1700 mOsm for a low range sample
–
The measured reading is above
4200 mOsm for a high range sample
Action:
•
Check for particulate matter in the sample.
•
Check the range setting currently being tested
(High Range or Low Range,) and confirm that
it is appropriate for the sample being tested.
•
Test a control such as Clinitrol.
Sample did
not freeze
Cause:
The sample might be above the range of
the instrument.
Action:
•
Confirm sample is present in the sampler.
•
.)
•
Clean the sample probe.
•
Test a control such as Clinitrol to see if the
issue persists.
Sample
pre-freeze
Cause:
The sample froze prematurely.
Action:
•
Re-run a new sample using a known control
such as Clinitrol.
•
Check for particulate matter in the sample.
•
Clean the sample probe with a chamber
cleaner dampened with water.
Test timeout
Cause:
The instrument was unable to complete
the test within the allotted time.
Action:
•
Check for particulate matter in the sample.
•
Test a control such as Clinitrol to see if the
issue persists.
•
If the issue persists, the sample might be
outside the range of the instrument.