Chapter 5
Configuring settings
OsmoTECH XT Single-Sample Micro-Osmometer Service Guide
30
Configuring OPC-UA communication
NOTE:
For more information about OPC-UA, see
1.
Press
NETWORKING
and log in.
2.
Press
OPC-UA
.
3.
Press
Enable.
4.
Enter the server port, device name, and minimum
sample interval (in msec).
5.
Press
SAVE
.
6.
Power off the instrument and wait 15 seconds for the
OPC-UA change to take affect.
7.
Power on the instrument.
Enabling the web server
Use the embedded web server to view and print
information—such as results, status, and data logs—from
a network computer or mobile device.
1.
Connect the instrument to the network using an
Ethernet cable.
2.
Press
NETWORKING
and log in.
3.
Press the
WEB SERVER
tab.
4.
Select
Web Server Enabled
.
5.
Press
to change the number of records (5-1000)
to display and press
SAVE
.
6.
Press and select
LAN
to display the web server IP
address.
7.
Enter the IP address into the address bar of any web
browser to display, save, and print the data.