
Chapter 5
Configuring settings
OsmoTECH XT Single-Sample Micro-Osmometer Service Guide
41
Resetting instrument system settings
To change the serial number, perform a factory reset,
and/or perform a system reset, complete the following
steps:
1.
Select
SYSTEM UPDATE
and log in to the Service
account.
2.
Press the
System Settings
tab.
3.
To change the serial number:
a.
Press the
Serial Number
tab.
b.
Press
, enter the serial number, and press
Enter
.
c.
Press
SAVE
.
4.
To reset the instrument to factory settings:
a.
Press the
Factory Reset
tab.
b.
Press
RESET
.
5.
To perform a system reset (factory reset and also
clears calibration, resistance, alignment, and serial
number):
a.
Press the
System Reset
tab.
b.
Press
RESET
.
Configuring V&V settings
To enable system screen captures, short system
timeouts, and disable calibration requirements,
complete the following steps:
1.
Select
SYSTEM UPDATE
and log in to the Service
account.
2.
Press the
V&V
tab.
3.
To enable short timeouts, select the
Use short
timeouts
checkbox.
4.
To bypass
Calibration required
errors and allow
service engineers to perform tests, select the
Ignore
Calibration Required
checkbox.
This checkbox is reset (cleared) when the
instrument is powered down and then powered up.
5.
Press
SAVE
.