Chapter 5
Configuring settings
OsmoTECH XT Single-Sample Micro-Osmometer Service Guide
28
5.
Enter a date and time format:
•
Select
Use Regional Format Settings t
o
automatically set the date and time formats based
on the language settings
(see
).
•
Otherwise, enter the
Date
,
Year
, and
Time
formats,
and the
Delimiter
as needed.
Entering the date and time manually
To enter the date and time manually (without
synchronizing to a server):
1.
Press
DATE & TIME
and log in.
2.
Clear the
Use Time Server
checkbox if it is selected.
3.
Enter the
Date
.
4.
Enter the
Time
.
5.
Select
Use Regional Format Settings
checkbox to
automatically set the date and time formats based on
the language settings
(see
)
and time format.
Otherwise, clear the
Use Regional Format Settings
checkbox and enter the
Date
,
Year
, and
Time
formats, and the
Delimiter
as needed.
6.
Press
SAVE
.
5.3 Changing the language
The language is set to English at the factory. You can
change it to one of nine other languages.
1.
Press
LANGUAGE
and log in.
2.
Select a language and press
SAVE
.
3.
Press
Exit
to return to the Settings menu.
NOTE:
If the system detects unsaved changes, it
prompts you to save the changes before
exiting.
4.
Power off the OsmoTECH XT and wait 15 seconds for
the language change to take affect.
5.
Power on the instrument.
5.4 Changing the probe resistance
Only
change the probe resistance setting after replacing
the sample probe or block probe.
After changing a probe resistance setting, you must
perform a thermal module setup and then recalibrate the
instrument.
1.
Press
PROBE RESISTANCE
and log in to open the
PROBE RESISTANCE
screen.