6240B DC Voltage Current Source/Monitor Operation Manual
6.7.3 Remote Command List
6-29
Sweep Random sweep
memory data
N [adr]
P
Random sweep memory data setting starts from the N
command and completes at the P command.
N<adr>,SVR<n>,SOV<data1>,SOV
<data2>,...,P (for voltage setting)
N<adr>,SIR<n>,SOI<data1>,SOI
<data2>,...,P (for current setting)
adr: Memory address (0 to 7999)
data1: Voltage or current source value at adr
data2: Voltage or current source value at
adr+1
0
0
*6
N? [adr]
Response: N<adr>,SVR<n>,SOV
<data>,P
(For voltage source value)
N<adr>,SIR<n>,SOI
<data>,P
(For current source value)
adr: <dddd>
n: <d>
data: <d.ddddE
d>
*1
NP?
Query of the random sweep memory setting status
Response: 0
...
Random sweep memory setting complete
1... Random sweep memory is being set
0
RSAV
Saves the random sweep data.
RLOD
Loads the random sweep data.
RCLR
Initializes the random sweep data. (Data saved in the
memory is not initialized.)
Pulse sweep base
value
BS [data]
data:
Pulse sweep base value
0
BS?
Response: BS ± <d.ddddE ± d>
*1
Bias value
SB [data]
data:
Bias value
0
SB?
Response: SB ± <d.ddddE ± d>
*1
RTB
(Return To Bias)
RB0
OFF
(Stays at the final value when sweep stops.)
RB1
ON
(Returns to the bias value when sweep
stops.)
RB?
Query of the RTB
Response: RB0 or RB1
*1:
The decimal point position in response varies depending on the setting value.
For the source value, limit value, and time parameter setting ranges, refer to the specifications.
*6:
Not initialized by the RINI or *RST command.
Item
Command
Description
Default
Operation
Power ON
Factory
shipment
DC/pulse
OPR/SUS
Sweep
OPR/SUS
Note:
1. When no source range is specified, the optimal range is
set.
2. Source values different from the present source function
cannot be set.