6240B DC Voltage Current Source/Monitor Operation Manual
5.2.10 Calculation Functions
5-57
5.2.10.4
Max/Min Calculation
1. Calculation expression
The Max/Min calculation obtains the maximum, minimum, average, and total values while the cal-
culation is set to ON.
2. Calculation result
Select the parameters,
H) COMPUTE
and
4) View Mx/Mn
on the menu screen to refer to the results.
1. Number of measurements
2. Maximum value
3. Minimum value
4. Average value
5. Total value
3. Operation
•
Valid data except over-range data and error data is calculated.
•
When the maximum or minimum value is updated with the notice buzzer set to ON, the buzzer
sounds. However, the buzzer may also sound when the displayed data does not change. This is
because the measurement resolution is smaller than the display resolution.
•
The Max/Min calculation is turned OFF by changing the measurement function or executing the
*RST command.
•
When the total value exceeds
999.999E+24, "Over" is displayed.
•
The calculation result is cleared and the calculation restarts under the following conditions:
1. Switching the NULL calculation between ON and OFF
2. Changing the NULL value
3. Switching the scaling calculation between ON and OFF
4. Changing any of the scaling constants