6240B DC Voltage Current Source/Monitor Operation Manual
5.2.3 Operations in Sweep Source Mode
5-14
2. Changing the sweep measurement parameters
The sweep measurement parameters are basically changeable only in Standby or Suspend status, but
the following parameters are changeable during sweep stop in Operate status.
•
Time parameter
•
Hold time
•
Source delay time
•
Measurement delay time
•
Pulse width
•
Period time
•
Start address and stop address for random sweep
(These are changeable in the start address to stop address range when switching from Standby
status to Operate or Suspend.)
•
Sweep function parameter
•
Repeat count
•
Reverse mode ON or OFF
•
RTB ON or OFF
•
Measurement auto range ON or OFF
•
Measurement ON or OFF
•
Measurement integration time
•
Selecting COMPLETE OUT or SYNC OUT external control signal output function
3. Indicator display for sweep status
indicates the sweep status.
Sweeping:
The indicator rotates.
HOLD:
The indicator stops rotation but remains displayed.
STOP:
The indicator is not displayed.