6240B DC Voltage Current Source/Monitor Operation Manual
8. CALIBRATION
8-1
8.
CALIBRATION
This chapter describes how to calibrate the 6240B to perform within the specified accuracy ranges.
In order to use the 6240B in the specified accuracies, periodic calibration at least once a year is recom-
mended.
Contact an ADC CORPORATION sales representative for the calibration service.
8.1
Measuring Instrument and Cables Necessary for Calibration
The following table shows the required accuracies of a DMM and cables used for calibration.
Range
ZERO
FS
Cable
Calibration
point
Required
accuracy
Calibration
point
Required
accuracy
300 mV
0 V
1.5
V
300 mV
15 ppm
A01044
(supplied accessory) *1
3 V
5
V
3 V
10 ppm
15 V
50
V
15 V
20 ppm
30
A
0 A
50 pA
30
A
120 ppm
300
A
500 pA
300
A
120 ppm
3 mA
5 nA
3 mA
120 ppm
30 mA
50 nA
30 mA
120 ppm
300 mA
500 nA
300 mA
210 ppm
1 A
5
A
1 A
170 ppm
4 A
5
A
1 A
170 ppm
*1:
When much externally induced noise exists, use shielded cables, such as A01001, etc.