6240B DC Voltage Current Source/Monitor Operation Manual
4.2.9 NULL/SEL Key
4-16
4.2.9
NULL/SEL Key
NULL key:
Switches the NULL calculation ON or OFF.
For more information on the NULL calculation function, refer to
Section 5.2.10.1, "NULL Calculation."
SEL Key:
Selects the parameters at the Input/Run level in the menu.
For more information, refer to Section 2.2.3, "Menu Operation."
4.2.10
OPR/SUSPEND Key (Operate/Suspend)
OPR key:
Switches between Operate and Suspend.
Operate
Turns the output status ON and the OPR indicator turns ON.
Displays measurement values when the measurement is ON. Not
display when the measurement is OFF. The operation varies
depending on the source mode.
DC:
Generates a setting value and displays the measurement
value.
PLS:
Generates a pulse and displays the measurement value.
DC-SWP or PLS SWP:
Generates a bias value and waits for measurement.
Produces sweep data and waits for trigger input. The
trigger input starts the sweep and displays a
measurement value.
PLC-RM:
Generates pulse current with different polarities and
displays the low-resistance measurement value.
Suspend
Outputs the suspend voltage without turning OFF the output
relays. The OPR indicator blinks. The suspend voltage is dis-
played and the header indicates the suspend status.
HZ:
High impedance status
LZ:
Low impedance status.
SUSPEND key (SHIFT):
Sets Suspend in Standby or Operate status.
NOTE:
Sweep data is produced when the output status is switched as follows in the DC sweep or pulse sweep
mode:
• Switching from Standby to Operate
• Switching from Standby to Suspend
• Switching to Operate after changing the sweep parameters during Suspend.