6240B DC Voltage Current Source/Monitor Operation Manual
6.7.3 Remote Command List
6-33
Cal
cula
tion
NULL calculation NL0
OFF
NL1
ON
NL?
Response: NL0 or NL1
KNL
data Sets the NULL constant. (An error occurs when the
NULL calculation is set to OFF.)
*4
0
KNL?
Response: KNL
d.dddddE
dd
Comparator
calculation
CO0
OFF
CO1
ON
CO?
Response: CO0 or CO1
KHI
data
Sets the upper limit.
0
KLO
data Sets the lower limit.
*4
0
KHI?
Response: KHI
d.dddddE
dd
KLO?
KLO
d.dddddE
dd
Scaling calcula-
tion
SCL0
OFF
SCL1
ON
SCL?
Response: SCL0 or SCL1
KA a
a: Constant A (0 (zero) is invalid.)
1
KB b
b: Constant B
0
KC c
c: Constant C
*4
1
KA?
Response: KA
d.dddddE
dd
KB?
KB
d.dddddE
dd
KC?
KC
d.dddddE
dd
MAX/MIN
calculation
MN0
OFF
MN1
ON
MN?
Response: MN0 or MN1
AVE?
Reads out the average value.
0
MAX?
Reads out the maximum value.
-9.26
MIN?
Reads out the minimum value.
+9.26
TOT?
Reads out the total value.
0
AVN?
Reads out the measurement count.
Response: AVN d.dd
0
*4:
The setting range is from 0 to
999.999E + 24.
Item
Command
Description
Default
Operation
Power ON
Factory
shipment
DC/pulse
OPR/SUS
Sweep
OPR/SUS