Chapter 4
Run the Relative Standard Curve Experiment
Start the Run
71
Applied Biosystems 7500/7500 Fast Real-Time PCR System Getting Started Guide for Relative Standard Curve
and Comparative C
T
Experiments
Notes
Run Guidelines
When you set up the 7500/7500 Fast system for automatic notification:
• Your system must be set up for network use. Refer to the
Applied Biosystems 7500/7500
Fast Real-Time PCR System Maintenance Guide
.
• Select the events for which you want to receive e-mail notifications. Select:
–
Instrument Error
– To notify recipients by email of all errors encountered by
the instrument during each run.
–
Run Started
– To notify recipients by email when the instrument starts a run.
–
Run Completed
– To notify recipients by email when the instrument completes a
run.
• Obtain e-mail addresses of those whom you want to be notified.
IMPORTANT!
Separate addresses with a comma (,).
• Contact your systems administrator or information technology department if you
need:
– The e-mail addresses of users who will receive notifications
– A network address for a simple mail transfer protocol (SMTP) server on the local
area network (LAN)
– A user name and password for the server, if required for access
– The Secure Sockets Layer (SSL) setting of the server (on or off)
Start the Run
IMPORTANT!
While the 7500/7500 Fast instrument is performing a run, do not create
experiments, perform maintenance, or allow the computer to run antivirus software or to
enter hibernation mode. Performing such activities while the instrument is running an
experiment will cause gaps in data collection.
To start your 7500/7500 Fast instrument:
1.
In the 7500 software, click
Run
in the navigation pane.
2.
Click
START RUN
.