Chapter 2
Design the Relative Standard Curve Experiment
Set Up the Samples
33
Applied Biosystems 7500/7500 Fast Real-Time PCR System Getting Started Guide for Relative Standard Curve
and Comparative C
T
Experiments
Notes
• Select which targets to test in the samples:
– Select
All Sample/Target Reactions
to test all targets in all samples.
– Select
Specify Sample/Target Reactions
to specify the targets to test in each
sample.
IMPORTANT!
When you use the Design Wizard to set up a relative standard curve
experiment, you can set up only singleplex reactions (amplification and detection of
one target per well). If you want to set up a relative standard curve experiment with
multiplex reactions (amplification and detection of two or more targets per well),
design your experiment using Advanced Setup instead of the Design Wizard.
Set Up Biological Replicate Groups
1.
Select
Specify Biological Replicate Groups
.
2.
Enter the number of biological replicate groups that you want to test in the reaction
plate.
3.
For each biological replicate group, click the cell in the Biological Group Name
column, then enter a name for the biological group.
4.
Assign the biological replicate group(s) to the reaction plate:
a.
Select wells in the plate layout that contain samples associated with the
biological replicate group.
b.
In the Assign column, select the check box of the appropriate biological
replicate group.
IMPORTANT!
A sample cannot belong to more than one biological group.
For More
Information
For more information on:
• Completing the Samples screen – Open the 7500 Software Help by clicking
or
pressing
F1
.
• Using Advanced Setup – See “Advanced Setup Workflow” on page 204.
1
2
4b
3