![abi 7500 Скачать руководство пользователя страница 182](http://html.mh-extra.com/html/abi/7500/7500_getting-started-manual_451420182.webp)
Chapter 9
Analyze the Comparative C
T
Experiment
Analyze the Experiment
Applied Biosystems 7500/7500 Fast Real-Time PCR System Getting Started Guide for Relative Standard Curve
and Comparative C
T
Experiments
158
Notes
Analyze the Experiment
The 7500 software analyzes the experiment and displays results in the analysis screens
(for example, the Amplification Plot screen, QC Summary screen, and so on).
About the
Example
Experiment
For the comparative C
T
example experiment, use the data file that is installed with the
7500 software. This data file was created with the same design parameters that are
provided in Chapter 6, then run and analyzed on a 7500/7500 Fast instrument.
The data file for the example experiment is on your computer at:
<drive>
:\Applied Biosystems\
<software name>
\experiments\
Comparative Ct Example.eds
where:
•
<drive>
is the computer hard drive on which the 7500 software is installed.
•
<software name>
is the current version of the 7500 software.
Analyze the
Example
Experiment
1.
Double-click
(7500
software) or select
Start
All Programs
Applied Biosystems
7500 Software
<software name>
where
<software name>
is the current version of the 7500 software.
2.
In the Home screen, click
Open
.
3.
In the Open dialog box, navigate to the
experiments
folder at:
<drive>
:\Applied Biosystems\
<software name>
\experiments\
4.
Double-click
Comparative Ct Example.eds
to open the example experiment data
file.
Note:
The examples folder contains several data files; be sure to select
Comparative Ct Example.eds
.