![abi 7500 Скачать руководство пользователя страница 127](http://html.mh-extra.com/html/abi/7500/7500_getting-started-manual_451420127.webp)
Chapter 5
Analyze the Relative Standard Curve Experiment
View the QC Summary
103
Applied Biosystems 7500/7500 Fast Real-Time PCR System Getting Started Guide for Relative Standard Curve
and Comparative C
T
Experiments
Notes
View the QC Summary
The QC Summary screen displays a list of the 7500 software flags, and includes the flag
frequency and location for the open experiment.
About the
Example
Experiment
In the standard curve example experiment, you review the QC Summary screen for any
flags generated by the experiment data. In the example experiment, the HIGHSD flags
has been generated for wells: A7, A8, and A9.
View the QC
Summary
1.
In the navigation pane, select
Analysis
QC Summary
.
Note:
If no data are displayed, click
Analyze
.
2.
Review the Flags Summary. In the example experiment, three wells are flagged.
3.
In the Flag Details table, look in the Frequency and Wells columns to determine
which flags appear in the experiment. In the example experiment, the Frequency
column displays 3 for the HIGHSD flag.
Note:
The Frequency column displays 0 to indicate that the flag does not appear in
the experiment.
4.
(Optional) Select the
HIGHSD
flag row to display details about the flag.
2
3