Chapter 6
Design the Comparative C
T
Experiment
Set Up the Samples
Applied Biosystems 7500/7500 Fast Real-Time PCR System Getting Started Guide for Relative Standard Curve
and Comparative C
T
Experiments
122
Notes
8.
In the Well Count pane, verify that there are:
• 96 Unknown wells
• 0 Negative Control wells
• 0 Empty wells
9.
In the Arrange Plate by drop-down list of the View Plate Layout tab, select
Rows
(default).
10.
Click
Next
.
Design
Guidelines
When you design your own comparative C
T
experiment:
• Identify each sample with a unique name and color. You can enter up to
100 characters in the Sample Name field.
• Enter the number of replicates (identical reactions) to set up. Applied Biosystems
recommends three replicates for each sample reaction.
• Set up biological replicate groups (see
“Set Up Biological Replicate Groups” on
page 123).
Biological replicates allow you to assess the representative nature of your results as
they relate to the population being studied. Inclusion of biological replicates can
give insight into any natural variation that is present within the population.
• The software automatically includes three negative controls for each target assay.
• Select which targets to test in the samples:
– Select
All Sample/Target Reactions
to test all targets in all samples.