![abi 7500 Скачать руководство пользователя страница 203](http://html.mh-extra.com/html/abi/7500/7500_getting-started-manual_451420203.webp)
Chapter 9
Analyze the Comparative C
T
Experiment
View the QC Summary
179
Applied Biosystems 7500/7500 Fast Real-Time PCR System Getting Started Guide for Relative Standard Curve
and Comparative C
T
Experiments
Notes
View the QC Summary
The QC Summary screen displays a list of the 7500 software flags, and includes the flag
frequency and location for the open experiment.
About the
Example
Experiment
In the comparative C
T
example experiment, you review the QC Summary screen for any
flags generated by the experiment data; however, no flags are generated.
View the QC
Summary
1.
In the navigation pane, select
Analysis
QC Summary
.
Note:
If no data are displayed, click
Analyze
.
2.
Review the Flags Summary. In the example experiment, there are no flagged wells.
3.
In the Flag Details table, look in the Frequency and Wells columns to determine the
flags that appear in the experiment. In the example experiment, no wells were
flagged by the 7500 software.
Note:
A 0 displayed in the Frequency column indicates that the flag does not appear
in the experiment.
4.
(Optional) Click each flag row to display details about the flag.
2
3
4