![abi 7500 Скачать руководство пользователя страница 113](http://html.mh-extra.com/html/abi/7500/7500_getting-started-manual_451420113.webp)
Chapter 5
Analyze the Relative Standard Curve Experiment
View the Amplification Plot
89
Applied Biosystems 7500/7500 Fast Real-Time PCR System Getting Started Guide for Relative Standard Curve
and Comparative C
T
Experiments
Notes
6.
Locate any outliers:
a.
In the Plot Type drop-down list, select
C
T
vs Well
.
b.
Look for outliers in the amplification plot. In the example experiment, there are
no outliers for GAPDH.
7.
Repeat steps 2 through 6 for the c-myc target. In the example experiment, there is
one outlier for c-myc (well A9). You omit this well in the troubleshooting section
(“Omit Wells from the Analysis” on page 105).
6a
outlier