Chapter 3
Prepare the Relative Standard Curve Reactions
Prepare the Standard Dilution Series
Applied Biosystems 7500/7500 Fast Real-Time PCR System Getting Started Guide for Relative Standard Curve
and Comparative C
T
Experiments
56
Notes
2.
Add the required volume of water (diluent) to each empty tube:
3.
In the c-myc Std. 1 tube:
a.
Vortex the stock for 3 to 5 seconds, then centrifuge the tube briefly.
b.
Using a new pipette tip, add 8.07 µL of stock to the c-myc Std. 1 tube.
c.
Vortex Std. 1 for 3 to 5 seconds, then centrifuge the tube briefly.
4.
In the c-myc Std. 2 tube:
a.
Using a new pipette tip, add 2.02 µL of dilution 1 to the c-myc Std. 2 tube.
b.
Vortex Std. 2 for 3 to 5 seconds, then centrifuge the tube briefly.
5.
In the c-myc Std. 3 tube:
a.
Using a new pipette tip, add 2.02 µL of dilution 2 to the c-myc Std. 3 tube.
b.
Vortex Std. 3 for 3 to 5 seconds, then centrifuge the tube briefly.
6.
In the c-myc Std. 4 tube:
a.
Using a new pipette tip, add 2.02 µL of dilution 3 to the c-myc Std. 4 tube.
b.
Vortex Std. 4 for 3 to 5 seconds, then centrifuge the tube briefly.
7.
In the c-myc Std. 5 tube:
a.
Using a new pipette tip, add 2.02 µL of dilution 4 to the c-myc Std. 5 tube.
b.
Vortex Std. 5 for 3 to 5 seconds, then centrifuge the tube briefly.
8.
Place the standards on ice until you prepare the reaction plate.
Prepare the
Standard Dilution
Series for the
GAPDH Assay
1.
Label a separate microcentrifuge tube for each standard:
• GAPDH Std. 1
• GAPDH Std. 2
• GAPDH Std. 3
• GAPDH Std. 4
• GAPDH Std. 5
Tube
Standard Name
Volume of Diluent to Add (µL)
1
c-myc Std. 1
12.10
2
c-myc Std. 2
18.15
3
c-myc Std. 3
18.15
4
c-myc Std. 4
18.15
5
c-myc Std. 5
18.15