Chapter 3
Prepare the Relative Standard Curve Reactions
Prepare the Reaction Plate
Applied Biosystems 7500/7500 Fast Real-Time PCR System Getting Started Guide for Relative Standard Curve
and Comparative C
T
Experiments
64
Notes
e.
While holding the film by the end-tabs, lower the film onto
the reaction plate (adhesive side facing the reaction plate).
Be sure the film completely covers all wells of the reaction
plate.
f.
While applying firm pressure, move the applicator slowly
across the film, horizontally and vertically, to ensure good
contact between the film and the entire surface of the
reaction plate.
g.
While using the applicator to hold the edge of the film in
place, grasp one end of the end-tab and pull up and away
sharply. Repeat for the other end-tab.
h.
Repeat step f to ensure a tight, evaporation-free seal. While
applying firm pressure, run the edge of the applicator
along all four sides of the outside border of the film.
Note:
Optical adhesive films do not adhere on contact. The films require the
application of pressure to ensure a tight seal.
i.
Inspect the reaction plate to be sure all wells are sealed. You should see an
imprint of all wells on the surface of the film.
For More
Information
For more information on:
• Preparing the reaction plate – Refer to the protocol appropriate for the reagents you
are using in the PCR reactions:
– TaqMan
®
Gene Expression Assays Protocol
– Custom TaqMan
®
Gene Expression Assays Protocol
• Consumables – See “Supported Consumables” on page 4.
• Using Advanced Setup to change the plate layout – See “Advanced Setup
Workflow” on page 204.