![abi 7500 Скачать руководство пользователя страница 106](http://html.mh-extra.com/html/abi/7500/7500_getting-started-manual_451420106.webp)
Chapter 5
Analyze the Relative Standard Curve Experiment
View the Standard Curve
Applied Biosystems 7500/7500 Fast Real-Time PCR System Getting Started Guide for Relative Standard Curve
and Comparative C
T
Experiments
82
Notes
View the Standard Curve
The Standard Curve screen displays the standard curve for samples designated as
standards. The 7500 software calculates the quantity of an unknown target from the
standard curve.
About the
Example
Experiment
In the relative standard curve example experiment, you review the Standard Curve screen
for the following values:
• Slope/amplification efficiency
• R
2
value (correlation coefficient)
• C
T
values
View the
Standard Curve
1.
In the navigation pane, select
Analysis
Standard Curve
.
Note:
If no data are displayed, click
Analyze
.
2.
Display all 96 wells in the Standard Curve screen by clicking the upper left corner of
the plate layout in the View Plate Layout tab.
3.
In the Plot Settings tab, select
All
in the Target drop-down list (default).
4.
In the Plot Settings tab, select
Default
in the Plot Color drop-down list.
5.
Click
(Show a legend for the plot).
Note:
This is a toggle button. When the legend is displayed, the button changes to
Hide the plot legend.
6.
View the values displayed below the standard curve. In the example experiment, the
values for each target are within the acceptable ranges:
Target
Slope
Y-Inter
R
2
Value
Amplification
Efficiency (Eff%)
GAPDH
-3.506
27.226
0.999
92.853
c-myc
-3.696
23.82
0.998
86.438