Chapter 9
Analyze the Comparative C
T
Experiment
View the Multicomponent Plot
Applied Biosystems 7500/7500 Fast Real-Time PCR System Getting Started Guide for Relative Standard Curve
and Comparative C
T
Experiments
184
Notes
Analysis
Guidelines
When you analyze your own comparative C
T
experiment, look for:
•
Passive reference
– The passive reference dye fluorescence level should remain
relatively constant throughout the PCR process.
•
Reporter dye
– The reporter dye fluorescence level should display a flat region
corresponding to the baseline, followed by a rapid rise in fluorescence as the
amplification proceeds.
•
Any irregularities in the signal
– There should not be any spikes, dips, and/or
sudden changes in the fluorescence.
•
Negative control wells
– There should not be any amplification in the negative
control wells.
For More
Information
For more information on the Multicomponent Plot screen, open the 7500 Software Help
by clicking
or pressing
F1
.