![abi 7500 Скачать руководство пользователя страница 93](http://html.mh-extra.com/html/abi/7500/7500_getting-started-manual_451420093.webp)
Chapter 4
Run the Relative Standard Curve Experiment
Enable the Notification Settings (Optional)
69
Applied Biosystems 7500/7500 Fast Real-Time PCR System Getting Started Guide for Relative Standard Curve
and Comparative C
T
Experiments
Notes
•
Reaction tubes
– Place the tubes in the PPH.
Note:
MicroAmp
®
Fast Reaction Tubes (PN 4358297) cannot be used in the
7500 Fast system.
IMPORTANT!
For optimal performance with partial loads:
For a 7500 Instrument
Load at least 16 tubes, arranging them first in the center columns of the
instrument system (columns 6 and 7). Move outward towards columns 1 and 12
as you add more tubes.
For a 7500 Fast Instrument
• Place empty tube strips in columns 1 and 12 to prevent crushing of tubes
containing samples.
• Place tube strips with samples in the PPH vertically, starting in columns 6
and 7 and moving outward.
• A maximum of 6 tube strips can be used in the 7500 Fast system. Leave
columns 2, 3, 10, and 11 empty.
4.
Close the tray door. Apply pressure to the right side of the tray and at an angle.
Enable the Notification Settings (Optional)
Enable the notification settings so that the 7500 software alerts you by email when the
7500/7500 Fast instrument begins and completes a run, or if an error occurs during a run.
Enabling the notifications settings is optional and does not affect the performance of the
7500/7500 Fast system or the duration of the run.
IMPORTANT!
The notification settings feature is available only if the computer that you are
using is running the 7500/7500 Fast instrument
and
is connected to an Ethernet network.
About the
Example
Experiment
In the example experiment:
• The 7500 software is set up to send notifications to three users (scientist, supervisor,
and technician at mycompany.com) when the 7500/7500 Fast system ends the run
and
if it encounters any errors during operation.
• The example SMTP server (smtp.mycompany.com) is set up for secure sockets layer
(SSL) encryption and requires authentication for use.
GR2471
GR2471
7500
tray top view
Push