ABEM Terrameter SAS 1000 / SAS 4000
- 84 -
33
Internal
error
-
invalid
setup
parameters
Try again. If the error persists, contact
ABEM.
34
Internal error - invalid fundamental
frequency
Try again. If the error persists, contact
ABEM.
41
Record full
No more readings can be added to this
record. Create a new record.
50
Low battery in ES 464
Change battery, or use an external power
supply
51
No response from ES 464 (LUND
Imaging System)
Check cables. Check ES 464 battery. Make
sure the "Serial" connector on the SAS
1000 / 4000 is not connected to any other
device.
52
No more items in LUND protocol file Check the protocol file
54
The specified protocol could not be
found
in
the
database
PROTOCOL.BIN
Check the protocol files. Recompile the
files and download again.
55
One or more addressed electrode
positions are outside the physical
layout
Check the electrode positions in the
protocol file.
56
Quit the protocol
57
Bad connection or no connection
during electrode test. One or more
electrodes.
Check
and
improve
the
electrode
grounding.
58
Bad connection or no connection at
all at remote current electrode
Check
and
improve
the
electrode
grounding.
59
No connection at any electrode
Check cables
60
Skip the rest of this protocol
61
Bad version of LUND protocol
The protocol read from disk is not
compatible with the program version.
Compile the protocol files with a version of
MPFC of same generation as the main SAS
1000 / 4000 program.
62
LUND connection error
Contact ABEM
63
No (or too few) ES464 units
connected
Check communication cables and the power
to each ES464 unit.
70
Multimac
logical
communication
error