ABEM Terrameter SAS 1000 / SAS 4000
- 75 -
9.6.3
Terrameter SAS LOG 300 - Checks
A full check of the SAS LOG 300 is quite complicated and must be done in the ABEM
laboratory. However, the thermometer and the fluid resistivity cell can be checked fairly
accurately using a fluid of known temperature and resistivity in a suitable container. Currently
ABEM advises against altering the factory set parameters for thermometer function.
To test the fluid resistivity cell, the entire logging probe must be submerged well below the
fluid surface. The return current terminal and the potential reference terminal must be well
separated from each other and also from the logging probe.