ABEM Terrameter SAS 1000 / SAS 4000
- 23 -
3.5
ACQUISITION TIMER
In all measurement modes (resistivity, IP and SP) and in as well standard measurements as
LUND measurements, a timer function can be activated. This timer will automatically start a
measurement at certain time intervals.
The Acquisition Timer is especially useful in monitoring applications. For example, you can
make Self Potential readings at certain time intervals, or take a complete LUND Imaging
layout each, say, four hours.
This is the standard measuring menu. When
you press
a measurement will be done. To
activate the acquisition timer, press .
Ch P1 P2 Resistance S.Dev Stacks .
1: .
2: N/A .
3: N/A .
4: N/A .
N=0 I=Auto .
Resistivity: Untitled, Untitled .
R00017.S4K C1= , C2= .
This brings up four commands. The
commands <Delete Last>, <Skip Next> and
<Exit> are obvious. If you select <Set Timer>
the next menu appears:
Ch P1 P2 Resistance S.Dev Stacks .
1:
╔══
Command
════╗
.
2: N/A
║
<Delete Last>
║
.
3: N/A
║
<Skip Next>
║
.
4: N/A
║
<Set Timer>
║
.
N=0
║
<Exit>
║
.
Resistivity:
╚═════════════╝
.
R00017.S4K C1= , C2= .
In the Set Timer menu you can define the start
time and date. The interval is the time in
hours, minutes and seconds between the start
of each measurement. If you define an
interval shorter than the time needed for SAS
1000 / 4000 to perform a measurement, the
next measurement will start immediately after
the previous.
The Set Timer menu has the same function
and
layout
in
the
case
of
LUND
measurements.
╔═════════════
Set Timer
════════════════╗
║
Start: 08/11/1999 13:30:00
║
║
Interval: 0000:00:01
║
║
║
║
║
║
║
║
║
╚
=Measure,
=Quit,
=Next/Prev Page
═╝