POL3L: 3-CHANNEL POLE-POLE CVES WITH 4 ELECTRODE
CABLES
Please observe that C2 and P2 are the remote electrodes.
This measurement protocol is designed for pole-pole CVES using four standard cable sections
linked together, addressing every fourth electrode take-out. The number of measurement
combinations included is less than half of the total possible number of independent
combinations using the pole-pole array (820 combinations), but designed to give an even roll-
along cover and reducing the measurement time.
Protocol
Array Type Address File
Size .ORG Size .UP/.DWN
POL2L
Pole-pole
LONG32
392
130
POL3L
Pole-Pole
LONG32
146
58
The POL3L protocol is identical to the POL2L, with the exception that it is designed for using
three measuring channels on the SAS 4000. Please note, that using this protocol the
measuring time is reduced by a factor of 2.2 to 2.7. In order to measure with three channels
you need the optional ”Pole-Pole” interconnection cable between the SAS 4000 and the ES
464, order no. 33 0020 14.
Figure 36. Data cover of 3 station roll-along for pole-pole CVES using POL3L (take-out
spacing 1metre).