ABEM Terrameter SAS 4000 / SAS 1000
- 128 -
Precision
V/I
Typically better than 0.5% (at 0.5 sec integration).
Dynamic range
Up to 140 dB plus 64 dB automatic gain
Induced Polarization (IP)
Number of input channels
4, galvanic isolated (SAS 1000 one channel)
Input impedance
10 M
, min
Automatic ranging
250 mV,
10 V,
400 V
Resolution
30 nV (theoretical), 1
V (at 0.5 sec integration time)
Type of IP measurements
Time domain chargeability (M) measured in
msec
Initial time delay
10 ms up to 10 sec in steps of 10 ms
Number of IP time intervals
1-10 user defined time windows on all channels.
Basic integration interval
20 msec or 16.67 msec depending on power line freq.
Dynamic range
Up to 140 dB plus 64 dB automatic gain
General
Computer
PC compatible
Memory capacity
> 1 000 000 data points
Temperature range, operation
-5
C to +50
C
Casing
Waterproof (IP66), rugged cast aluminum
Display
LCD, 200x64 pixels, 8 lines of 40 characters
Multi connector
Current and potential for all four channels. RS232
serial communication with PC or external devices.
Support for external devices
LUND Imaging System, SAS LOG 300, Multimac
Power
An ordinary car battery is recommended. NiMH clip-
on power pack (del. with SAS 4000) could be used in
low current soundings.
Power consumption
600 mA in standby. Up to 20 A during transmission
of high currents.
Weight
4.6 kg. (SAS EBC 1.2 kg, NiMH Power Pack 2.0 kg)
Dimensions
105 x 325 x 300 mm (W x L x H) with power pack