ABEM Terrameter SAS 1000 / SAS 4000
- 103 -
14.11
POL3SS: 3-CHANNEL POLE-POLE CVES WITH 2 ELECTRODE
CABLES
Please observe that C2 and P2 are the remote electrodes.
This measurement protocol is designed for pole-pole CVES using only two standard cable
sections, addressing each electrode take-out. The number of measurement combinations
included is only a fraction of the total possible number of independent combinations using the
pole-pole array (820 combinations), but designed to give an even roll-along cover and
reducing the measurement time.
The POL3SS protocol is designed for using three measuring channels using SAS4000 and
ES464. In order to measure with three channels you need the optional ”Pole-Pole”
interconnection cable between the SAS4000 and the ES464, order no. 33 0020 14.
Protocol
Array Type Address File
Size .ORG Size .UP/.DWN
POL3SS
Pole-pole
SHORT
297(106)
180(74)
Figure 35. Data cover of 3 station roll-along for pole-pole CVES using POL3SS (take-out
spacing 1metre).