ABEM Terrameter SAS 1000 / SAS 4000
- 73 -
9.6
TESTING OF LUND EQUIPMENT
9.6.1
Fast continuity test of cable and relay switch
It is possible to test each of the 32 conductors in a LUND cable by use of the ordinary LUND
program module, using a special protocol file.
1.
Connect both ends of one LUND cable to connector 1 and 2 on the ES464 /
ES10-64[C][e] as illustrated in the above figure.
2.
Connect the SAS 1000 / 4000 to the ES464 / ES10-64[C][e] as usual and turn on the
equipment.
3.
Start the “LUND Imaging System”, select Resistivity Mode and define a record.
4.
Select the protocol denoted “CABLETST”. If this protocol is not loaded in the instrument,
please compile a new PROTOCOL.BIN (containing this protocol) using the SAS 4000
Utility Software and download it into the SAS 1000 / 4000.
5.
Select the current 1000 mA, and Acq. delay and Acq. time = 0.1 sec. Select stacks=1 and
NO electrode test.
6.
Press MEASURE to start the measuring sequence.
7.
During the measuring sequence, the display will show the resistance in Ohm of each
conductor.
Figure 22.
Connecting both ends of one LUND cable to the ES464 / ES10-64(e).