188
System Manual SCK Gas Sampling System
42/23-81 EN Rev. 4
Step Procedure
2
Turn the T-handle
(1)
of the filter detaching device counter-
clockwise. This pulls the filter insert over the detaching disk
(3)
out
of the casing
(5)
.
3
Turn bridge
(2)
until it can be pulled off from the bridge holder
through the elongated holes.
4
Pull filter insert with bridge and detaching disk out.
5
Turn detaching disk until it can be pulled off from the hexagon
screws
(4)
over the elongated holes.
6
If necessary, mechanically clean the filter stone.
WARNING!
The filter insert may be soiled with toxic and corrosive
substances depending on the substances with which it comes into
contact in operation.
Wear appropriate protective clothing when
cleaning.
7
Replace O-rings
(6)
.
Note:
The O-rings do not need to be re-greased.
8
Reassemble filter insert and push back into casing, tighten T-
handle.
9
Unlock EMERGENCY STOP switch, turn service switch back to
OFF
position and insert probe again using the controller.
WARNING!
Probe moves into the kiln.
Do not stand in the area of
travel of the probe or probe carriage until the probe is com-
pletely inserted into the kiln and the warning lamps go out.
Содержание SCK
Страница 2: ......
Страница 16: ......
Страница 36: ......
Страница 54: ......
Страница 110: ......
Страница 222: ......
Страница 232: ......
Страница 234: ...234 System Manual SCK Gas Sampling System 42 23 81 EN Rev 4 Pneumatic hydraulic interface plan ...
Страница 235: ...42 23 81 EN Rev 4 System Manual SCK Gas Sampling System 235 ...
Страница 236: ...236 System Manual SCK Gas Sampling System 42 23 81 EN Rev 4 Electrical connection plan ...
Страница 237: ...42 23 81 EN Rev 4 System Manual SCK Gas Sampling System 237 ...
Страница 238: ...238 System Manual SCK Gas Sampling System 42 23 81 EN Rev 4 Transformer connection plan ...
Страница 242: ......
Страница 243: ......