186
System Manual SCK Gas Sampling System
42/23-81 EN Rev. 4
Step Procedure
10
Take out the filter stone.
WARNING!
The filter insert may be soiled with toxic and corrosive
substances depending on the substances with which it comes into
contact in operation.
Wear appropriate protective clothing when
cleaning.
11
Replace filter stone.
12
If necessary, replace seals (see
Gas sampling probe 60S: Replac-
ing the seal set
on page 187).
13
Reassemble filter insert and push back into casing, tighten T-
handle.
14
Unlock EMERGENCY STOP switch, turn service switch back to
OFF
position and insert probe again using the controller.
WARNING!
Probe moves into the kiln.
Do not stand in the area of
travel of the probe or probe carriage until the probe is com-
pletely inserted into the kiln and the warning lamps go out.