42/23-81 EN Rev. 4
System Manual SCK Gas Sampling System
217
Error
display
Plain text (not stored
in controller)
Cause Measures
Error Com-
pressed Air
Emergency
Tank
Minimum compressed-
air emergency tank
pressure not adequate
If the central compressed-air
supply is also faulty, first
remedy the cause for failure
of the central compressed-
air supply
Leak in tank supply pipes or
tank itself
Pressure switch 578 error
Check central compressed-air
supply
Check supply pipe to com-
pressed-air emergency tank
for leaks
Check functioning and setting
of pressure switch 578
Sample Gas
Probe Pres-
sure MIN
Sample gas main
alarm low pressure not
reached
Gas sampling probe clogged
Sample gas line clogged
Check and, if necessary, clean
gas sampling probe
Check and, if necessary, clean
sample gas line
Probe Heat-
ing Temp.
MIN
Probe heating temper-
ature too low
At start-up probe heating not
switched on long enough
Probe heating defective
Defect in temperature con-
troller
Heating
connection
cable
damaged
Heating sleeve defective
Wait for heating-up time (ap-
prox. 20 min)
Check probe heating for prop-
er functioning
Probe Heat-
ing Temp.
MAX
Probe heating temper-
ature too high
Probe heating overheating
Check PT100 sensor
Probe Heat-
ing Temp.
MIN
Probe heating temper-
ature too low
At start-up probe heating not
switched on long enough
Probe heating defective
Defect in temperature con-
troller
Heating
connection
cable
damaged
Heating sleeve defective
Wait for heating-up time (ap-
prox. 20 min)
Check probe heating for prop-
er functioning
Cable Drag
Chain Sam-
ple Gas Line
Temp. Error
Cable drag chain
sample gas line temp.
error
Cable drag chain heating not
switched on long enough
Cable drag chain heating
defective
Wait for heating-up time (ap-
prox. 20 min)
Check heating for proper
functioning
Error Posi-
tion Switch
Probe Action
Probe movement limit
switch error
Front and at least one of the
two rear limit switches for
probe movement triggered
simultaneously
Check function and position of
limit switches
Check position of trigger ele-
ment
Error 3x
Cleaning
Not Suc-
cessful
Probe H
Probe cleaning failed
after 3x cleaning
Cleaning cycle can no long-
er clean probe H adequately
Remove dirt sticking to the gas
inlet
Clean the filter element
Check compressed-air supply
and solenoid valves of probe
cleaning system
Error Posi-
tion Switch
Plunger
Action
Plunger movement
limit switch error
Both plunger movement limit
switches triggered simulta-
neously
Check function and position of
limit switches
Содержание SCK
Страница 2: ......
Страница 16: ......
Страница 36: ......
Страница 54: ......
Страница 110: ......
Страница 222: ......
Страница 232: ......
Страница 234: ...234 System Manual SCK Gas Sampling System 42 23 81 EN Rev 4 Pneumatic hydraulic interface plan ...
Страница 235: ...42 23 81 EN Rev 4 System Manual SCK Gas Sampling System 235 ...
Страница 236: ...236 System Manual SCK Gas Sampling System 42 23 81 EN Rev 4 Electrical connection plan ...
Страница 237: ...42 23 81 EN Rev 4 System Manual SCK Gas Sampling System 237 ...
Страница 238: ...238 System Manual SCK Gas Sampling System 42 23 81 EN Rev 4 Transformer connection plan ...
Страница 242: ......
Страница 243: ......