CI/AZ40-EN Rev. A
5
1
Overview / system dimensions
The Endura AZ40 combustion gas analyzer continuously
samples and analyzes combustion waste gases for both
oxygen and carbon monoxide equivalent (COe).
The analyzer has 4 main assemblies:
— transmitter (controller / display unit), see Fig. 2
— sensor assembly (housing a zirconia-based oxygen sensor,
catalytic COe sensor and an air powered aspirator),
see Fig. 3
— smart sensor electronics (part of the sensor assembly),
see Fig. 3
— probe / filter assembly, see Fig. 3
The analyzer uses a close-coupled sampling system where the
sensor assembly is mounted directly against the process wall.
The sample is filtered and drawn through the sensor assembly
by the air powered aspirator.
This combination of a short sample path and pumped sample
provides a very rapid response to changing gas concentrations.
The gas sample is held above the sample dew point to provide
analysis on a 'wet' basis and prevent acid gases from
condensing in the sample path.Thermocouple inputs for
process temperature measurement enable calculations of
combustion efficiency.
Fig. 2 Endura AZ40 transmitter – main components / dimensions
Fig. 3 Endura AZ40 sensor assembly – main components / dimensions
Endura AZ40 Gas Analyser
85 - 240v ~AC, 50/60Hz, 730W Max
TYPE No: AZ40/XXXXXXXXXXXXXXX
SERIAL No: 3K62000000000
TAG No:
DATE OF MANUFACTURE: W25/15
TRANSMITTER
Enclosure rating: IP66/NEMA4X
Entry Type:
Communication:
54
(2.1)
270
(10.6)
300
(11.8)
300 (11.8)
153 (6)
136 (5.4)
46.5
(1.8)
46.5
(1.8)
54
(2.1)
40
(1.6)
84 (3.3)
53 (2.1)
58 (2.3)
330 (13)
Cable entries – mains power in/sensor signal input and user I/O
Zero gas inlet
Span gas inlet
Calibration gas outlet
M20 or
1
/
2
in.
NPT entry
M25 or
3
/
4
in. NPT entry
M20 or
1
/
2
in.
NPT entry
367
(14.4)
92.4 (3.6)
127 (5)
154.5 (6.1)
141 (5.6)
Probe / Filter assembly
Smart sensor electronics
Cable entries –
mains power in / signal outputs
Sensor data label
Instrument air inlet
Calibration gas inlet
Vacuum test port
Standard 2 in. ANSI mounting adapter