CI/AZ40-EN Rev. A
27
(
Zero Gas Cal. and Span Gas Cal. Time
values are dependent
upon the length of pipe runs and the proximity
of the transmitter to the sensor. Default time = 10 mins.)
12.Press the
key (below the
Edit
prompt) and use the
/
keys to enter the required
Zero Gas Cal. Time
in
minutes.
When the required time is displayed, press the
key (below
the
OK
prompt) to set the value and display the
Span Gas
Cal. Time
screen:
13.Press the
key (below the
Edit
prompt) and use the
/
keys to enter the required
Span Gas Cal. Time
in
minutes.
When the required time is displayed, press the
key (below
the
OK
prompt) to set the value and display the
Recovery
Time
screen:
(
Recovery Time
is the time delay before the new sensor
value becomes live to the process.)
14.Press the
key (below the
Edit
prompt) and use the
/
keys to enter the required
Recovery Time
in
seconds.
When the required time is displayed, press the
key (below
the
OK
prompt) to set the value and display the
Calibrate
menu options screen:
15.Press the
key (below the
Back
prompt) twice to exit the
Calibrate
level.
16. Proceed to Section 7, page 28 to perform a calibration
routine.
Zero Gas Cal. Time
10 Mins
Back
Edit
Calibration Setup
10
Mins
Max
60
Min
1
Next
OK
Zero Gas Cal. Time
Span Gas Cal. Time
10 Mins
Back
Edit
Calibration Setup
10
Mins
Max
60
Min
1
Next
OK
Span Gas Cal. Time
Recovery Time
60 Secs
Back
Edit
Calibration Setup
060
Secs
Max
600
Min
1
Next
OK
Recovery Time
Back
Select
Calibrate
Manual Calibration
Manual Blowback
Calibration Setup
Blowback Setup
Scheduled Events
Flow Rate Test