26
CI/AZ40-EN Rev. A
7. Use the
/
keys to scroll to the
CO Zero Test Gas
menu and press the
key (below the
Select
prompt).
The
Calibration Setup / CO Zero Test Gas
screen is
displayed:
8. Press the
key (below the
Edit
prompt) and use the
/
keys to enter the required
CO Zero Test Gas
value.
When the required value is displayed, press the
key
(below the
OK
prompt) to set the value and display the
CO
Span Test Gas
screen:
9. Press the
key (below the
Edit
prompt) and use the
/
keys to enter the required
CO Span Test Gas
value.
When the required value is displayed, press the
key
(below the
OK
prompt) to set the value and display the
O2
Zero Test Gas
screen:
10.Press the
key (below the
Edit
prompt) and use the
/
keys to enter the required
O2 Zero Test Gas
value.
When the required value is displayed, press the
key
(below the
OK
prompt) to set the value and display the
O2
Span Test Gas
screen:
11.Press the
key (below the
Edit
prompt) and use the
/
keys to enter the required
O2 Span Test Gas
value.
When the required value is displayed, press the
key
(below the
OK
prompt) to set the value and display the
Zero
Gas Cal. Time
screen:
CO Zero Test Gas
0 ppm
Back
Edit
Calibration Setup
00000
ppm
Max
2000
Min
0
Next
OK
CO Zero Test Gas
CO Span Test Gas
2000 ppm
Back
Edit
Calibration Setup
20000
ppm
Max
2000
Min
0
Next
OK
CO Span Test Gas
O2 Zero Test Gas
10.00 % O2
Back
Edit
Calibration Setup
10.00
% O2
Max
25.00
Min
0.10
Next
OK
O2 Zero Test Gas
O2 Span Test Gas
25.00 % O2
Back
Edit
Calibration Setup
25.00
% O2
Max
25.00
Min
0.10
Next
OK
O2 Span Test Gas